• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Blurred Vision (2137); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Race: unknown/asked but not provided.The device has not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Health effect - clinical code: (b)(4).An attempt was made to obtain missing information; however, the account did not have patient race information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patient's operative eye due to blurry distance vision (dv) as the patient was unable to read street signs.The patient outcome: 20/40 -2, visual acuity (va) pre-operative (op): 20/30 -1, and va post-op: 20/40 -2.The iol was decentered during a post-op visit.Manifest refraction (mrx) +0.25-0.50x35 20/50.Surgeon tried to reposition the iol, but had to remove and replace iol.The incision was enlarged, however, there was no vitrectomy or sutures required.A monofocal lens was implanted as a replacement.Through follow-up, it was learned that the right eye was affected.A non-johnson & johnson lens was successfully implanted as a replacement.The patient is doing well when discharged.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14313718
Report Number3012236936-2022-01204
Device Sequence Number20090242
Product Code HQL
UDI-Device Identifier05050474750753
UDI-Public(01)05050474750753(17)240523
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU240
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Initial Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight64 KG
Patient EthnicityNon Hispanic
-
-