The customer reported a false negative dat (direct anti-globulin test) result of a proficiency sample with ahg anti-igg c3d.The customer stated that the reaction was supposed to be positive due to the igg sensitization.The customer provided the complaint sample anti-human globulin (ahg) of the supposedly defective lot for investigational testing and also the proficiency sample.Our quality control laboratory tested the proficiency sample with the complaint sample according to the instruction for use (immediate spin tube test) and could confirm the customer´s finding.The direct antiglobulin test (dat) was negative.The dat was also performed with prolonged incubation times of 5 and 15 minutes which are not according the instruction for use (ifu).Again, the dat was negative.Additionally, the proficiency sample was tested for dat in the ih-card ahg anti-igg and ih-card ahg anti-igg, -c3d.In both cards the dat was clearly positive.The igg sensitization of the proficiency sample was confirmed with monospecific anti-sera (anti-igg, anti-iga, anti-igm, anti-c3c and anti-c3d).Furthermore, the proficiency sample was tested again with the complaint sample in parallel with the ahg anti-igg and the ahg anti-igg, -c3d from ortho.In the immediate spin tube test the ahg from ortho showed a +/- reaction.In the tube test with an incubation for 5 minutes, the ahg anti-igg showed a 1+ positive reaction, while the complaint sample and the ahg from ortho showed a +/- reaction.All reactions not specifically mentioned were negative.The complaint sample and our retention sample of the affected lot were tested for potency and specificity.The testing was done in parallel with a reference lot.All acceptance criteria were met.All three reagents showed comparable results.Based on the investigation the complaint was classified as confirmed - epp (expected product performance).The complaint and the retention sample of the supposedly defective lot met the specifications regarding potency and specificity with anti-d sensitized red blood cells.Both ahg reagents showed negative reactions with the proficiency sample, but also the ahg from ortho did not show a positive reaction, but only a +/-.The positive reactions of the proficiency sample in the gel cards could be explained by the higher sensitivity of this method.In our tests, the age of the sample may have had a negative impact.In case of a weak sensitization at the border line a decrease in reactivity by aging might have caused the false negative test results.A weakly configured proficiency sample could also be responsible for the false negative result at the customer's site.The instructions for use (ifu) contain a relevant information: "negative reactions will be obtained if the sample contains antibodies present in concentrations too low to be detected by the test method employed.No test method is capable of detecting all red cell antibodies.".We were able to rule out a general problem with the ahg anti-igg, -c3d; polyspecific.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.
|