MAUDE Adverse Event Report: ANGIODYNAMICS VENACURE NEVERTOUCH DIRECT PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS VENACURE NEVERTOUCH DIRECT PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number 11403101
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device has yet to be returned to the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
Ansm (france national competent authority) notified angiodynamics of an event that occurred with a laser fiber from a never touch direct procedure kit.The end user had reported that during a procedure, the tip of the fiber became loose.The fiber was exchanged for a new of the same device to complete the procedure.There was no report of the patient experiencing any adverse effects, harm, or require medical intervention as a result of this incident.
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint of the fiber tip was loose could not be confirmed; no fiber sample was returned.Without receiving the complaint sample for evaluation a root cause for this event cannot be determined.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.During the manufacturing process, the disposable fiber device receives a 100% inspection and an aql inspection in which the quality of the fiber strip is inspected.Prior to packaging, all components are inspected for damage.Labeling review: the instructions for use which is supplied to the end user with this catalog number contains the following statement "avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a diameter of 16cm." the user manual, which is supplied to the end user with this unit, states "before using a fiber, check it carefully for any signs of damage during storage or transit.Protective caps should be in place over sma connectors.Do not use if there is any sign of damage." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Event or Problem Description
Additional information provided reported that the gold tip of the fiber fully detached outside the patient.
 
Additional Manufacturer Narrative
Reference (b)(4).
 
Additional Manufacturer Narrative
Returned for evaluation was an evlt nevertouch direct fiber and detached gold tip.It appeared that a fracture in the fiber glass core allowed laser energy to be diverted from the fiber glass core and compromised the bonds designed to hold the gold tip tube in place.Most likely the gold tip tube / quartz ferrule bond initially failed and then subsequently the crimp / adhesive bond between the gold tip tube and buffer was compromised which allowed the gold tip tube to move slightly during the initial procedure.Between the fractured fiber glass core and compromised tip tube bond, handling of the tip tube after the initial procedure was sufficient for complete separation of the gold tip tube / quartz ferrule to occur.The customer's reported complaint of the gold tip detached was confirmed.The most likely root cause for the tip detachment was handling of the tip tube after the initial procedure.Although the complaint description was confirmed, a definitive root cause could not be determined.It is possibly due to the fracture of the fiber glass core that compromised the bonds holding gold tip in place, or the result of an impact force against the tip area of the fiber assembly.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.During the manufacturing process, the disposable fiber device receives a 100% inspection and an aql inspection in which the quality of the fiber strip is inspected.Prior to packaging, all components are inspected for damage.Labeling review: the instructions for use which is supplied to the end user with this catalog number contains the following statement "avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a diameter of 16cm." the user manual, which is supplied to the end user with this unit, states "before using a fiber, check it carefully for any signs of damage during storage or transit.Protective caps should be in place over sma connectors.Do not use if there is any sign of damage." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENACURE NEVERTOUCH DIRECT PROCEDURE KIT
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key14315005
Report Number1319211-2022-00020
Device Sequence Number5000628
Product Code GEX
UDI-Device IdentifierH787114031015
UDI-PublicH787114031015
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K171921
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Catalogue Number11403101
Device Lot Number5705687
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/26/2022
Supplement Date Received by Manufacturer05/13/2022
06/08/2022
06/28/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date05/20/2022
06/09/2022
06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age18 YR
Patient SexMale
Patient Weight77 KG
-
-