Model Number 164267
Medical Device Problem Codes Failure to Power Up (1476); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The field service representative (fsr) verified that the centrifugal control unit (ccu) would shut down.After consulting with the manufacturer's technical support specialist on the log data, he replaced the centrifugal motor.The unit operated to the manufacturer's specification.The centrifugal motor involved was part number: 816572, serial number: (b)(4), udi: (b)(4).
 
Event or Problem Description
After use of the device for a cardiopulmonary bypass (cpb) procedure , the user reported that the centrifugal control unit (ccu) gave an 'overspeed' alarm when barely moving the speed control knob, then it would not power up at all.There were no reported adverse consequences to the patient.
 
Additional Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) started the centrifugal pump up and a 'stopped: overspeed' message appeared.When the pst opened the pump up, liquid dribbled out.The pst observed that the magnets and screws were thoroughly rusted and there were exposed wires and contacts.It was determined that the centrifugal motor did not meet specification.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key14315453
Report Number1828100-2022-00184
Device Sequence Number20045398
Product Code DTQ
UDI-Device Identifier00886799000458
UDI-Public(01)00886799000458(11)170727
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K172220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number164267
Device Catalogue Number164267
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date06/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-