| Model Number |
RLT231418 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to gore that a gore® excluder® aaa endoprosthesis (rlt231418), which was intended to be used for an endovascular aneurysm repair (evar), would not advance through the gore® dryseal flex introducer sheath (dsf1633).An alternative gore® excluder® aaa endoprosthesis (rlt231416) was used instead and the case was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
No patient specific details have been provided.Therefore, the patient initials reflect the w.L.Gore internal case number.A review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.The device evaluation concluded on april 25, 2022 showed the following: visual examination of the device showed the lock pin was dislodged from the olive.Engineering attempted to insert the device through a 16 fr introducer sheath per ifu but was unable to do so.Based on the findings from this evaluation, the physician¿s observation that the device would not advance through the introducer sheath was confirmed.The likely cause for the device unable to would advance in the introducer sheath was unable to be determined with the available information.The likely cause for the dislodged lock pin is forcibly inserting the device into the introducer sheath.Per the gore® excluder® aaa endoprosthesis instructions for use (ifu), trunk-ipsilateral leg endoprosthesis positioning warning states: do not continue advancing any portion of the delivery system if resistance is felt during advancement of the guidewire, sheath, or catheter.Stop and assess the cause of resistance.Vessel or catheter damage or premature deployment have occurred and may result in potential patient harms, see adverse events.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: added health effect - impact code 2199 - no health consequences or impact.Added type of investigation codes 4111 and 4119.Added investigation findings code 213 - no device problem found.Investigation findings code 3243 refer to the type of investigation code 10, with investigation conclusions codes 61 and 22.Investigation findings code 213 refers to the type of investigation code 3331, with investigation conclusions code 4315.Code 4111 and 4119: multiple attempts have been made to acquire additional information to classify any possible specific device problem, such as reason for the advancement issue, patient details, and details of usage of the introducer sheath.No additional information has been provided.
|
| |
|
Search Alerts/Recalls
|