| Model Number |
N/A |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of refx0144 showed seven other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported "ir placed 4f single lumen picc.Clinician reported leaking from the end cap.End cap was replaced a few times.A new sl picc was placed.The same thing occurred.They then tried a different end cap (needless connector) and this resolved the issue." this report addresses the first device.
|
| |
|
Event or Problem Description
|
|
It was reported "ir placed 4f single lumen picc.Clinician reported leaking from the end cap.End cap was replaced a few times.A new sl picc was placed.The same thing occurred.They then tried a different end cap (needless connector) and this resolved the issue." this report addresses the first device.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling based on a review of this information, the following was concluded: the complaint of a leak at the luer hub is confirmed to be manufacturing related.One 4 fr powerpicc catheter was returned for evaluation.The catheter was returned trimmed at the 1 cm depth marker.The catheter was infused with water and the sample was observed to leak at the hub connection.Microscopic inspection of the purple luer adapter revealed a protrusion in the luer material within the orifice.The mold defect appeared to prevent firm attachment of the injection cap and luer lock syringe; therefore, the complaint is confirmed.Bd is working closely with the manufacturing facility to prevent reoccurrence of the reported event.
|
| |
|