• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT 5FDL REVERSE TAPER PICC BASIC TRAY W/MICROINTRODUCER 45CM; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 GROSHONG NXT 5FDL REVERSE TAPER PICC BASIC TRAY W/MICROINTRODUCER 45CM; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of rees3128 showed one other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that "groshong picc was inserted in radiology.Patient was sent back to surgical ward after insertion.Ward medical officer was about to start treatment for this patient.While he was checking patency, he found that there was leaking at the proximal hub.He stopped the procedure.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a catheter leak is confirmed but the exact cause remains unknown.One video sample of a 5 fr groshong dl catheter was provided for evaluation.The catheter is shown within patient use and is secured with a statlock securement device.The catheter is being manipulated to bend the catheter at the region that extends from the bifurcation hub.A spraying leak is noted at the proximal portion of the catheter where the interface is located with the hub.The characteristics of the damage on the catheter could not be closely inspected from the video sample provided.Since the presence of a leak was observed in the video provided, the complaint is confirmed but the exact cause remains unknown.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported that "groshong picc was inserted in radiology.Patient was sent back to surgical ward after insertion.Ward medical officer was about to start treatment for this patient.While he was checking patency, he found that there was leaking at the proximal hub.He stopped the procedure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GROSHONG NXT 5FDL REVERSE TAPER PICC BASIC TRAY W/MICROINTRODUCER 45CM
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14316610
Report Number3006260740-2022-01662
Device Sequence Number5224069
Product Code LJS
UDI-Device Identifier00801741035388
UDI-Public(01)00801741035388
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K023374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number7857505
Device Lot NumberREES3128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-