| Model Number |
0502537030 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: eib trevor, rep.Poor visibility.(b)(4).The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be a dislodged or broken rod lens inside the scope causing the image of the scope to become blurry and have poor visibility.This may have occurred as a result of rough handling during product transport and/or sterilization the product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurryimage.
|
| |
|