| Model Number |
7171Q/58 |
| Medical Device Problem Codes |
Break (1069); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented in clinic for follow up.Upon review, the right ventricular lead exhibited noise leading to over-sensing.The patient presented to a scheduled explant procedure.During the procedure, while removing the lead, the tip electrode was damaged.The lead was explanted and replaced.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of tip electrode separation was confirmed while the reported event of noise was not confirmed.As received, a partial lead (distal portion) was returned in one piece with all four tines intact.Visual inspection of the lead found that the inner coil/distal tip was pulled out/stretched from the insulation which is consistent with procedural damage.The cause of reported event of tip electrode separation was isolated to procedural damage.Electrical test and x-ray examination were normal.
|
| |
|