• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL SDN. BHD. HUDSON HUMID-VENT FLEX TUBE; SET, TUBING AND SUPPORT, VENTI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL SDN. BHD. HUDSON HUMID-VENT FLEX TUBE; SET, TUBING AND SUPPORT, VENTI Back to Search Results
Model Number IPN043837
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the material of the devices is tearing open.Multiple attempts for additional information was requested to the customer has been unsuccessful.Patient condition unknown at time of report.
 
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
It was reported that "the material of the devices is tearing open".Multiple attempts for additional information was requested to the customer has been unsuccessful.Patient condition unknown at time of report.
 
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as the customer has reported that the sample is not available for investigation; however, the customer did provide a photo.The photo was reviewed and it was confirmed that the tubing was torn.The manufacturing site reports: based on the observation, this complaint is confirmed.The defect part was supplied by our supplier.In manufacturing, a non-conformance (nc) has been raised to address this issue and root causes with corrective actions will be implemented through this nc process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON HUMID-VENT FLEX TUBE
Common Device Name
SET, TUBING AND SUPPORT, VENTI
Manufacturer (Section D)
TELEFLEX MEDICAL SDN. BHD.
perak, west malaysia
Manufacturer (Section G)
TELEFLEX MEDICAL SDN. BHD.
lot no : pt2577 jalen perusahaan
4 kamunting industrial estate
perak, west malaysia 34600
MY   34600
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key14317865
Report Number8040412-2022-00109
Device Sequence Number5270742
Product Code BZO
UDI-Device Identifier04026704388769
UDI-Public04026704388769
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN043837
Device Catalogue Number20011
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer05/27/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-