• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIVA (TM) ID TRACTIP HIGH POWER SINGLE-USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIVA (TM) ID TRACTIP HIGH POWER SINGLE-USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006R8403961
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Approximated based on the date the manufacturer became aware of the event.This event was reported by a bsc sale rep.The reported healthcare facility is: (b)(6).(b)(4).Investigation results: the returned flexiva id tractip high power single-use laser fiber was analyzed, and a visual evaluation noted that the laser fiber was returned in one piece.The fiber body measured 315 cm from the distal tip to the proximal end of the sma connector.The exposed glass tip measured 4 mm and had normal degradation.A functional evaluation revealed that the light from the sma connector was distorted, indicating that there was fracture within the connector.When tested on the laser console, it read error #67: fiber expired! connect new fiber and resume operation.No other problems with the device were noted.The reported event of the flexiva id tractip high power single-use laser fiber stopped working was confirmed.The reported laser fiber stopped working is consistent with a fiber broken within the connector.It is likely that procedural handling of the device during set-up or use contributed to the fracture within the fiber connector.A broken fiber within the connector of the device can result in insufficient aiming beam and low laser energy output, up to a complete inability for the fiber to fire.Review and analysis of all available information indicated that the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2022 that flexiva id tractip high power single-use laser fiber was used in a ureteroscopy procedure in the ureter performed on an unknown date.During the procedure, the laser fiber stopped working.There were no patient complications reported as a result of this event.Note: this event has been deemed reportable based on the investigation finding that the laser fiber was broken within the connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA (TM) ID TRACTIP HIGH POWER SINGLE-USE LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
COHERENT INCORPORATED
5100 patrick henry drive
santa clara CA 95054
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14317883
Report Number3005099803-2022-02567
Device Sequence Number4661087
Product Code GEX
UDI-Device Identifier08714729963844
UDI-Public08714729963844
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K110685
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/12/2024
Device Model NumberM006R8403961
Device Catalogue Number69270
Device Lot Number0000010920
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/15/2022
Initial Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-