• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 3 HI OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 3 HI OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-14-030
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Discomfort (2330); Embolism/Embolus (4438); Joint Laxity (4526); Unspecified Musculoskeletal problem (4535); Swelling/ Edema (4577)
Date of Event 10/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Claim letter received.Patient was revised due to failed total right hip replacement.Both surgeries were performed at hospital for special surgery in nyc.Doi: (b)(6) 2019.Dor: (b)(6) 2021.Right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned for evaluation.The investigation could not draw any conclusions about the reported event without the device to examine.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
After review of medical records, it was reported that patient complained of squeaking, sharp pain, and discomfort.The patient was then revised, operative notes reported that there was black viscous fluid consistent with metal debris.There was notable metal transfer in a striped pattern on the femoral head.The poly liner was removed and the wear pattern was consistent with impingement in an extended and externally rotated femoral position.Patient taking eliquis for a month to prevent blood clots as she had suffered blood clot form initial surgery.The squeaking noise and pain was from the femoral head rubbing directly against the acetabular shell.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 3 HI OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14318202
Report Number1818910-2022-08414
Device Sequence Number4516836
Product Code KWA
UDI-Device Identifier10603295001058
UDI-Public10603295001058
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012-14-030
Device Catalogue Number101214030
Device Lot NumberJ42J47
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2022
Supplement Date Received by Manufacturer05/05/2022
05/26/2022
06/24/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date05/13/2022
06/07/2022
06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/02/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX +4 NEUT 32IDX48OD; APEX HOLE ELIM POSITIVE STOP; DELTA CER HEAD 12/14 32MM +9; PINN 100 W/GRIPTION 48MM; TRI-LOCK BPS SZ 3 HI OFFSET
Patient Age48 YR
Patient SexFemale
-
-