| Catalog Number |
MZ5307 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: unknown.Medical device expiration date: unknown.Device manufacture date: unknown.Medical device lot #: 21096362.Medical device expiration date: 29-sep-2024.Device manufacture date: 28-sep-2021.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd maxzero multi-fuse pressure rated extension set with needleless connector(s) experienced a case of leakage, and at least one case of component separation.The following information was provided by the initial reporter: during product training, the nurse reports that when transferring a patient from the stretcher to the bed, the maxzero extension cord became entangled at the time of transfer and broke.With no damage impact to the user, the device was replaced.The care management team states that the extenders break easily at the connection of the bifurcation to the lumens.
|
| |
|
Event or Problem Description
|
|
It was reported that bd maxzero multi-fuse pressure rated extension set with needleless connector(s) experienced a case of leakage, and at least one case of component separation.The following information was provided by the initial reporter: during product training, the nurse reports that when transferring a patient from the stretcher to the bed, the maxzero extension cord became entangled at the time of transfer and broke.With no damage impact to the user, the device was replaced.The care management team states that the extenders break easily at the connection of the bifurcation to the lumens.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to corrected information: d.9.Device available for eval?: no.D.9.Returned to manufacturer on: n/a.H.6.Investigation: a mz5307 product was not available for investigation; however the customer confirmed that the complaint sample was from lot 21096362.As part of the investigation, the customer provided a video of a mz5307 product; analysis of the video identified that the set was being pulled apart with force by the user.A review of the production records for lot 21096362 did not identify any in-process testing failures or quality deviations which may have resulted in a report of this nature.A definitive root cause for the customer's experience could not be determined, however the customer confirmed that the product had separated after becoming tangled during a patient transfer.As part of the feedback the customer provided a video which showed the customer intentionally pulling apart an unused infusion set; please note the products are manufactured inline with the requirements of bs: en: iso 8536-9: infusion equipment for medical use.Fluid lines for use with pressure infusion equipment.The standard states that the products are manufactured to withstand 15n of force over a 15 second period; the video sent by the customer suggests that the product was subjected to far in excess of this force.H3 other text : see h.10.
|
| |
|