• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMBU A/S AMBU VIVASIGHT 2 DLT; DOUBLE-LUMEN ENDOBRONCHIAL TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMBU A/S AMBU VIVASIGHT 2 DLT; DOUBLE-LUMEN ENDOBRONCHIAL TUBE Back to Search Results
Model Number 412371000
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During the procedure with vivasight 2 dlt the cuff malfunctions were identified shortly after intubation and prior to initiation of the surgical procedure.The tube was replaced with a second tube.The patient had to be re-intubated with a third tube to gain control the airway during the procedure.It was confirmed that there was no harm to the patient and no impact on patient outcome.Two replacements and a total of three dlt tubes are required to gain control of the airway during the operation.
 
Additional Manufacturer Narrative
The investigation was performed on retention samples with same manufacturing lot#10000575876.Tracheal cuff functionality was checked as per working instruction.To simulate the condition prior the intubation, the tracheal cuffs were inflated, and no leakages were observed.Further simulation was performed by over-inflating the tracheal cuffs and hold them in tight condition (to simulate the condition in the tracheal area).Then, movement was introduced to the tubes to observe for possible cuff leakage during patient movement.Based on the simulation result, we were able to see the similar failure mode as per reported failure.The cuffs leaked (damage) due to high friction between the lumen wall and over inflated cuff surfaces.The customer has not provided any information on whether a pre-check was performed prior to the procedure in accordance to the instruction for use.The instruction for use states that the user should test the cuff for integrity by completely inflating and deflating prior to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMBU VIVASIGHT 2 DLT
Common Device Name
DOUBLE-LUMEN ENDOBRONCHIAL TUBE
Manufacturer (Section D)
AMBU A/S
baltorpbakken 13
ballterup, dk 2750
DA  2750
MDR Report Key14319551
Report Number1220828-2022-00006
Device Sequence Number17448358
Product Code CBI
UDI-Device Identifier05707480145737
UDI-Public05707480145737
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/25/2024
Device Model Number412371000
Device Catalogue Number412371000
Device Lot Number1000575876
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/12/2022
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date05/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-