| Catalog Number |
309.48 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient age reported as in the (b)(6).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6), 2022, the patient underwent removal surgery and open reduction internal fixation with the trochanteric fixation nail advanced (tfna) system for a fracture of the subtrochanteric femur which occurred after the patient¿s initial surgery.During the surgery, it was confirmed that the head of two screws were broken; the surgeon tried to remove the screws with a hollow reamer.The tip of the reamer tube broke.The surgeon used another reamer tube; however, this reamer tube also broke.The surgeon managed to remove the broad plate and the narrow plate and tried to fix the tfna long nail.When the surgeon tried to drill the bone with a non-synthes intramedullary nail drill for insertion of distal locking screw, the tip of the intramedullary nail drill was deformed by the hardness of the bone.The surgeon managed to drill and make a screw hole.During insertion of the locking screw, the tip of the inter-lock screwdriver deformed and could not be used.The surgery was completed successfully with less than thirty minutes delay.Patient was stable.This report is for a spare reamer tube for hollow reamer (309.450).This is report 4 of 4 for (b)(4) and captures the intraoperative events that occurred on (b)(6), 2022.The postoperative subtrochanteric fracture is captured under (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9 h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that sparereamertube f/309.450 was found to be broken, the broken part was received in the evidence provided.A dimensional inspection for the sparereamertube f/309.450 was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the sparereamertube f/309.450 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: part: 5787 (309.480) lot: 3l19901 manufacturing site: bettlach release to warehouse date: 01 april 2019 a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|