GLOBUS MEDICAL, INC. ANTHEM DISTAL RADIUS FRACTURE SYSTEM; ANTHEM VOLAR PLATE, DOUBLE ROW, STANDARD, LEFT, 3 HOLE, TI Back to Search Results
Model Number 1171.1313
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported there was a revision for screws backing out of the left anthem plate 10 days post operatively.
 
Additional Manufacturer Narrative
The device was returned for evaluation.Initial observation showed discoloration and wear around and above the swept cuts of the polyaxial hole.The imaging provided shows screws backed out very far out of the volar plate.The patient suffered bilateral wrist fractures; both wrists were operated on and both saw the same screw back out issues in the head of the plate.Additional information provided that a poly cone was used for drilling instead of a torque limiter.Screws not inserted to the proper torque could contribute to a screw back out.It's also possible that the screws were inserted outside the ±20° cone of angulation; however, an exact cause to the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM DISTAL RADIUS FRACTURE SYSTEM
Common Device Name
ANTHEM VOLAR PLATE, DOUBLE ROW, STANDARD, LEFT, 3 HOLE, TI
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key14321186
Report Number3004142400-2022-00070
Device Sequence Number5936012
Product Code HWC
UDI-Device Identifier00889095390384
UDI-Public00889095390384
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K163361
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1171.1313
Device Lot NumberBAY402FH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/08/2022
Initial Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-