Model Number 8801-90034
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tip of an ozark screwdriver fractured during intra-operative screw insertion.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.
 
Additional Manufacturer Narrative
The device was visually inspected and the threaded inner tip of the driver is fractured off.Distal hexalobe of driver is slightly deformed rotationally.A review of complaint history associated with the subject catalog and lot number was performed, and no adverse trends were observed.It was further reported via follow up communication that the reported event occurred during screw insertion.The blue spin top handle was used and excessive force was not applied.Patient had hard bone.Ozark surgical technique guide was reviewed: ¿the threaded driver has a threaded distal tip to aid in the retention of the screws during insertion.Attach the proximal end of the threaded driver to the torque limiting handle.Insert the distal end of the threaded driver into the screw head and turn the knob clockwise to engage screw.¿ note: ¿the threaded driver must be used with the 12 in-lbs torque limiting handle.¿.The most likely cause of the reported event was determined to be due to not using the 12 in-lb torque limiting handle with the threaded driver.Torque limiting driver will prevent the tip of the driver from experiencing too much torque (over 12 in-lbs) which can damage the driver.The patients hard bone quality may have contributed to the event by causing the driver to be over torqued without a torque limiting handle.
 
Event or Problem Description
It was reported that the tip of an ozark screwdriver fractured during intra-operative screw insertion.There were no adverse consequences to the patient.The procedure was completed successfully without surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW DRIVER W/ THREAD; SIZE 10
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14322055
Report Number3004774118-2022-00178
Device Sequence Number11927318
Product Code LXH
UDI-Device Identifier10888857400863
UDI-Public10888857400863
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8801-90034
Device Catalogue Number8801-90034
Device Lot NumberTCFU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/08/2022
Supplement Date Received by Manufacturer06/30/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-