COOK INC FLEXOR RAABE GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G11636
Medical Device Problem Codes Material Separation (1562); Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluated by mfg: device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, during a femoral angiogram of the right leg, the middle of a flexor raabe guiding sheath unraveled and separated in the patient.Scarring and plaque were reported at the left groin access site.After the sheath separated, the physician obtained access lower in the leg and used an unspecified balloon device to push the sheath up to retrieve it.The sheath was removed in its entirety.The dilator was not in place at the time of separation.The procedure was completed.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Summary of event: as reported, during a femoral angiogram of the right leg, the middle of a flexor raabe guiding sheath unraveled and separated in the patient.Scarring and plaque were reported at the left groin access site.After the sheath separated, the physician obtained access lower in the leg and used an unspecified balloon device to push the sheath up to retrieve it.The sheath was removed in its entirety.The dilator was not in place at the time of separation.The procedure was completed.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The sheath was separated, exposing the inner coil, with the distal section measuring approximately 16.3cm and the proximal section measuring approximately 52.2cm.Both sections were also accordioned.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu warns that reinsertion of the dilator prior to removal of the sheath ¿increases the strength of the sheath and lessens the risk of device separation¿ and suggests insertion of the dilator prior to removal if resistance is encountered or anticipated during withdrawal of the device.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that although it is possible that the patient¿s scarred anatomy and plaque, as well as unintended user error related to failure to reinsert the dilator contributed to the event, a definitive cause for this event cannot be established at this time.There is currently a capa investigation open to further investigate this failure mode.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information has been received since the last report was submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR RAABE GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14322792
Report Number1820334-2022-00695
Device Sequence Number5220061
Product Code DYB
UDI-Device Identifier00827002116368
UDI-Public(01)00827002116368(17)250208(10)14521117
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/08/2025
Device Model NumberG11636
Device Catalogue NumberKCFW-6.0-38-70-RB-RAABE
Device Lot Number14521117
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/22/2022
Supplement Date Received by Manufacturer08/04/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexMale
Patient Weight68 KG
-
-