• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV SYSTEM M.SA 10 U.SPRUNG RES.; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV SYSTEM M.SA 10 U.SPRUNG RES.; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV425T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 03/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav shuntsystem (part # fv425t) was implanted during a procedure performed on an unknown date.According to the complainant, the physician suspected a drainage issue.The patient underwent a revision procedure on an unknown date.The complaint device has not been returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.
 
Event or Problem Description
Patient details: age: 56 years.Weight: 75 kilograms (kg).Height: 173 centimeters (cm).Gender: unknown.
 
Additional Manufacturer Narrative
Investigation: visual inspection: the following observations were made during the visual inspection: - slightly visible deformation.- coloured distal catheter.Measurement of surface of the housing: no deformation detected, measured value: -0,002 mm [tolerance (0 ± 0.02mm)].Permeability test: the test showed that the progav shunt system is non-permeable.To determine the location of the occlusion, the shunt system was dismantled and each element was tested individually for permeability.The test showed that the peritoneal catheter is difficult permeable, while the progav shunt system without the catheter is permeable.Computer control test: the computer controlled test showed the opening pressure of the progav, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 9,02 cmh2o.This is within the specified tolerance of 10 cmh2o ± 3 cmh2o.Additionally, the paedishuntassistant was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 10 cmh2o in the vertical position, a pressure of 10 cmh2o +4/-2 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the paedishuntassistant had a pressure of 6,34 cmh2o.This is not within the specified tolerance of 10 cmh2o +4/-2 cmh2o.According to our results, we can detect a presence of an accelerated outflow.Adjustability test: the progav was found to be not adjustable to specifications.The paedishuntassistant is a fixed pressure valve, therefore the adjustability test is inapplicable.Braking force and brake function test: the braking force test indicates that the brake function of the progav is present.Due to the non-adjustability of the valve, an investigation of the braking force was not possible.Internal inspection of product: in order to verify whether the investigated shunt system was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valves are finally opened with a special tool.After dismantling of the valves, deposits were found in both valves.Results: based on our investigation results, we have determined that there was an accelerated outflow and a non- adjustability in the shuntsystem at the time of our investigation.Detected deposits were the cause of the accelerated outflow and the non- adjustaility.Existing deposits in the csf can temporarily impair the function of the valve and is described as concomitant symptoms in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV SYSTEM M.SA 10 U.SPRUNG RES.
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key14323036
Report Number3004721439-2022-00125
Device Sequence Number11908869
Product Code JXG
UDI-Device Identifier04041906132224
UDI-Public4041906132224
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/26/2023
Device Model NumberFV425T
Device Catalogue NumberFV425T
Device Lot Number20040351
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexUnknown
Patient Weight75 KG
-
-