BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6531
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The patient presented with coronary artery disease.The target lesion was located in the left anterior descending artery.A 12mm x 2.50mm nc quantum apex balloon catheter was advanced for pre-dilation.However, it was noted that the shaft of the balloon broke while advancing into the lesion.The physician removed the whole system out and after that took an alternate access for completing the procedure.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
Returned product consisted of an incomplete device return of an nc quantum apex balloon catheter.The device was visually and microscopically examined.There was a separation of the shaft, 117.2cm from the strain relief.A total of 11.7cm of inflation lumen was returned and the full returned length of the lumen was stretched down.The guidewire lumen, balloon, markerbands, and tip failed to return for analysis.Product analysis confirmed the reported break as there was separation of the shaft to the device.
 
Event or Problem Description
It was reported that shaft break occurred.The patient presented with coronary artery disease.The target lesion was located in the left anterior descending artery.A 12mm x 2.50mm nc quantum apex balloon catheter was advanced for pre-dilation.However, it was noted that the shaft of the balloon broke while advancing into the lesion.The physician removed the whole system out and after that took an alternate access for completing the procedure.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14323656
Report Number2134265-2022-04602
Device Sequence Number4586884
Product Code LOX
UDI-Device Identifier08714729783244
UDI-Public08714729783244
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/24/2025
Device Model Number6531
Device Catalogue Number6531
Device Lot Number0028740921
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/14/2022
Supplement Date Received by Manufacturer07/21/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date08/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-