• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ULNAR COMPONENT PLASMA SPRAYED SIZE 5 75 MM LENGTH LEFT FOR CEMENTED USE ONLY; PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ULNAR COMPONENT PLASMA SPRAYED SIZE 5 75 MM LENGTH LEFT FOR CEMENTED USE ONLY; PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED Back to Search Results
Catalog Number 00840001507
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870); Failure of Implant (1924); Osteopenia/ Osteoporosis (2651)
Date of Event 08/12/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an initial implantation approximately three (3) years ago as part of a clinical study.Subsequently, patient had fallen and suffered a peri-prosthetic fracture near the humeral component.Patient was not revised and was being treated conservatively with a splint.At the two (2) year follow up, the site reported loosening of the humeral component at the site of the peri-prosthetic fracture.Examination of x-rays also noted loosening of ulnar component.Currently, the site notes this as mild and tolerated with no current plan for revision.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Item# 00840004510; lot# 63691823, item# 00840009500; lot# 63930799.Full establishment name : (b)(6).Foreign: (b)(6) customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.H6-component code recommendation mechanical (g04) - stem.The following sections were updated/corrected: b4; b5; g3; g6; h1; h2; h3; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: two views of the left elbow demonstrate a left total elbow arthroplasty with loosening of the proximal ulnar component.Mild radiolucency along the bone cement interface of the distal humeral component could indicate loosening as well.No bony fracture seen.Radial head resection.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULNAR COMPONENT PLASMA SPRAYED SIZE 5 75 MM LENGTH LEFT FOR CEMENTED USE ONLY
Common Device Name
PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14324628
Report Number0001822565-2022-01346
Device Sequence Number4891216
Product Code JDC
UDI-Device Identifier00889024271241
UDI-Public(01)00889024271241(17)270831(10)63670155
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K181307
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00840001507
Device Lot Number63670155
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer05/12/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date05/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/11/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRAITVE IN H10.
Outcome Attributed to Adverse Event Other;
Patient Age74 YR
Patient SexFemale
Patient Weight58 KG
-
-