Model Number ROB10025
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a cori demo the real intelligence tracking camera had a soft bleep, and the led was red.The real intelligence tracking camera did not pass the initial test.As this happened in a demonstration, there was not patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Section h10: the real intelligence tracking camera (netherlands) rob10025, (b)(6), intended for use in treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was completed.The reported problem was confirmed.The camera was connected to a cori system.The front right led light was flashing red with an audible alert.The shock sensor was reset per procedure and the camera functions properly.The most likely cause of this event is the power source (battery) to the shock sensor circuit has disconnected itself from the shock sensor circuit during camera transport resulting in the front camera leds blinking red and an audible beep.Refer to the cori surgical system user¿s manual for proper handling.The robotics cart and accessories should be moved carefully over thresholds and bumps to avoid system damage.Exercise care during the surgical site setup, surgery and tear down to protect the device from an unforeseen force, such as falling onto a hard surface or damage.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar prior events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE TRACKING CAMERA
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14324700
Report Number3010266064-2022-00346
Device Sequence Number4697823
Product Code OLO
UDI-Device Identifier00885556757437
UDI-Public00885556757437
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10025
Device Catalogue NumberROB10025
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2022
Supplement Date Received by Manufacturer05/25/2022
07/25/2022
Initial Report FDA Received Date05/07/2022
Supplement Report FDA Received Date05/26/2022
07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORI ROBOTICS USA: ROB20000, SN: (B)(6).
Patient SexUnknown
-
-