| Model Number |
ROB10025 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Failure to Calibrate (2440)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a cori demo the real intelligence tracking camera had a soft bleep, and the led was red.The real intelligence tracking camera did not pass the initial test.As this happened in a demonstration, there was not patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: the real intelligence tracking camera (netherlands) rob10025, (b)(6), intended for use in treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was completed.The reported problem was confirmed.The camera was connected to a cori system.The front right led light was flashing red with an audible alert.The shock sensor was reset per procedure and the camera functions properly.The most likely cause of this event is the power source (battery) to the shock sensor circuit has disconnected itself from the shock sensor circuit during camera transport resulting in the front camera leds blinking red and an audible beep.Refer to the cori surgical system user¿s manual for proper handling.The robotics cart and accessories should be moved carefully over thresholds and bumps to avoid system damage.Exercise care during the surgical site setup, surgery and tear down to protect the device from an unforeseen force, such as falling onto a hard surface or damage.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar prior events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.Internal complaint reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|