OLYMPUS MEDICAL SYSTEMS CORP. VISERA PRO XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S40PRO
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2022
Event Type  malfunction  
Manufacturer Narrative
In addition, service found the output connector (light guide connection) was loose.The chassis and the top cover were corroded.The high intensity mode was not working.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported the subject device would not power up.The reported issue occurred while preparing the subject device for use during an unspecified procedure.There was no patient or user injury reported due to the event.The device was returned to an olympus service center for evaluation.During inspection and testing, service confirmed the device would not power up due to a power unit failure.In addition, service found the fuse box had burn marks and the ac power inlet was burned.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device manufacture date is unknown, therefore, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the legal manufacturer's investigation, the phenomenon was reproduced.It was presumed that the event occurred due to a converter failure.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA PRO XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14326532
MDR Text Key299779920
Report Number8010047-2022-07849
Device Sequence Number1
Product Code GCT
UDI-Device Identifier04953170228933
UDI-Public04953170228933
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S40PRO
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2022
Was the Report Sent to FDA? No
Date Manufacturer Received05/18/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-