|
Model Number 420142-04 |
Device Problems
Thermal Decomposition of Device (1071); Arcing (2583)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Intuitive surgical, inc.(isi) received the monopolar cautery involved with this complaint and completed the device evaluation.Failure analysis investigations confirmed the reported issue.The instrument was found to have thermal damage to its grip tips and broken electrical contacts at the distal tip.The instrument failed the electrical continuity test.Functional testing of energy delivery could not be performed due to the thermal damage.While the root cause for this failure is typically attributed to the user, it is unknown what caused the damage for this event.Additional finding not related to the reported thermal damage: the instrument was found to have a dislodged flush tube inside the housing.Root cause of this failure is typically attributed to the user.A review of the instrument log for the monopolar cautery (part # 420142-04, lot # s10131218 983) associated with this event has been performed.The instrument was last used on (b)(6) 2020 on system (b)(4).The alleged event occurred on the 14th use.There is no indication that the instrument was used in subsequent procedures after the alleged event reported on this record.No image or video clip for the reported event was submitted for review.This complaint is being reported due to the following conclusion: the monopolar cautery instrument tip was found melted and charred.Melting/charring is evidence of the deliverance of electrical current at a location other than intended.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
|
|
Event Description
|
It was reported that during central processing, the monopolar cautery instrument tip was found melted and charred.It was removed from service and replaced.There was no reported injury.Intuitive surgical, inc.(isi) obtained the following information about the complaint: there was nothing abnormal at the visual inspection on the sterile field.No arcing was observed.The monopolar cord was not connected to a bipolar instrument.The customer used conmed system 5000 generator with settings: cut: 40, coag: 40, bip 35.The monopolar scissors were used during 90% of case with the maryland bipolar forceps.Often during the dissection, the tips were covered in carbonized tissue requiring being removed and cleaned.There was no patient harm and the surgery completed as planned.
|
|
Search Alerts/Recalls
|
|
|