• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON INFINITI SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON INFINITI SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752088
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample has been received by manufacturing that has not yet been evaluated.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during a cataract surgery metallic powder-like substance was left on the iris during surgery.Surgeon suspect foreign object was from the phacoemulsification handpiece, phaco tip, sleeve, or pack.There are no plans for surgery to remove the foreign material and the patient is scheduled for observation.
 
Additional Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A piece of metal was reported.A wet intrepid plus fluidics management system (fms) was returned and was visually inspected.The collected fluid from the cassette and drain bag was filtered and separated through a filter.The filter was placed into petri dishes and sent to particle lab for further evaluation.Microscopic examination showed a reflective particle.The reflective particle was isolated and analyzed.Analysis identified titanium, and vanadium.The presence of these elements along with the visual characteristics suggest the part is a titanium alloy.The filters were examined with the hitachi scanning electron microscope (sem) equipped with an edax energy dispersive x-ray spectrometer (eds).Reflective particles, dark particles and orange particles were found.The sem/edax analysis could not conclusively identify any further materials.A review of the consumable components indicated the cassette and tubing manifolds are not manufactured with titanium alloy and would not have originated from a part manufactured at this manufacturing site.The particles would not have generated from the manufacturing process.The root cause of the customers complaint could not be conclusively determine the source of the titanium alloy.Titanium alloy is not material purchased or used in the manufacturing process.No action will be taken as manufacturer does not manufacture a component with titanium.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14332430
Report Number1644019-2022-00346
Device Sequence Number4587182
Product Code LRO
UDI-Device Identifier00380657520886
UDI-Public00380657520886
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Catalogue Number8065752088
Device Lot Number2480199H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/12/2022
Supplement Date Received by Manufacturer05/09/2022
Initial Report FDA Received Date05/08/2022
Supplement Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HANDPIECE TIP.; INFINITI VISION SYSTEM.; OZIL TORSIONAL HANDPIECE.
Patient SexUnknown
-
-