| Model Number |
21-7024-24 |
| Medical Device Problem Code |
Accessory Incompatible (1004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on a picture received and one product returned.The complainant returned units were visually inspected.The sample has a flash in the square arch of the housing; thus, the failure mode reported is confirmed.The molding tool was repaired, and a picture of the defect was added for operators to be aware of and inspect for the condition.A device history review was conducted which indicated all inspections were completed and root cause analysis was conducted by supplier.
|
| |
|
Event or Problem Description
|
|
It was reported that there was difficulty in inserting cassettes onto pump tried several pumps without success.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
While performing a review of submitted mdr report it was discovered that this file was inadvertently assessed as reportable.No patient death, serious injury, and no evidence of a reportable product malfunction.The hazardous situation for the given complaint device would not likely cause or contribute to a death or serious injury if the malfunction were to recur.Report 3012307300-2022-08134 is no longer considered reportable, please disregard any mdr reports associated with it.
|
| |
|