(b)(4).Batch #: v94742.Investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nslg2c45a device was returned with the bottom portion of the i blade out of the lower jaw channel; the lower jaw channel was damaged.In addition, device was received with skiving of the heat shrink (shaft cover) material.All of the heat shrink material appeared to have been returned.The device was connected to the generator and it was recognized.Because of the condition of the device not all functional testing could be performed with the generator.The jaw was unable to cycle open and close.A probable cause of the damage to the jaw could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.It should be noted that product failure is multifactorial.The ifu warns "prior to inserting or removing the device from the trocar, ensure that the device is in the straight, non-articulated position by confirming that the indicator arrow on the articulation wheel is aligned with the top of the instrument." the ifu also warns the user to discontinue use if black heat shrink covering is damaged.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's related to the reported complaint condition were identified.
|