| Model Number |
QRP8 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of the photographic evidence, showed dried organic matter (since cleaned, but presumed to be dried blood) running in a horizontal thin line around one of the rollers, we do not have enough evidence to know exactly how long that organic matter had been present, but from the photographs we can see that it was most likely to have been present before the second tube was inserted and would certainly contribute to the very quick degradation of the second tubing set installed.We have found nothing significant with the product to suggest it was faulty equipment.
|
| |
|
Event or Problem Description
|
|
Tubing rupture on arterial roller during pediatric case.Perfusionist changed out raceway and placed new raceway into roller.After approximately 10 minutes the new tubing could be seen wearing away in the raceway.Roller pump subsequently changed out.Looking at roller pump pictures there appears to be a defect in one of the rollers at the base which is circled in black.
|
| |
|
Additional Manufacturer Narrative
|
|
No root cause found.
|
| |
|
Event or Problem Description
|
|
Tubing rupture on arterial roller during pediatric case.Perfusionist changed out raceway and placed new raceway into roller.After approximately 10 minutes the new tubing could be seen wearing away in the raceway.Roller pump subsequently changed out.Looking at roller pump pictures there appears to be a defect in one of the rollers at the base which is circled in black.
|
| |
|