MAUDE Adverse Event Report: LDR MÉDICAL ROI-A MEDIAN IMPLANT PXL : 23X26 H 12MM 10 DEGREES; ROI-A ALIF CAGE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MÉDICAL ROI-A MEDIAN IMPLANT PXL : 23X26 H 12MM 10 DEGREES; ROI-A ALIF CAGE SYSTEM Back to Search Results
Catalog Number IR2433P
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that upon malleting the implant into the patient's lumbar spine at l5-s1, the cage cracked and a piece broke at the point of attachment in the area of the threaded rod; the fractured piece was recovered.The implant and first plate were removed and new ones were placed to complete the procedure without patient impacts.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that upon malleting the implant into the patient's lumbar spine at l5-s1, the cage cracked and a piece broke at the point of attachment in the area of the threaded rod; the fractured piece was recovered.The implant and first plate were removed and new ones were placed to complete the procedure without patient impacts.
 
Additional Manufacturer Narrative
Corrections in d4: udi number.Additional information in d4: expiration date, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation product was not returned and photos were not provided.Device evaluation could not be completed.Potential cause root cause was unable to be determined.This event could possibly be attributed to not using the starter awl or inserting the plates off-axis.Dhr review per dhr review, the part was likely conforming when it left zimvie control.Device use this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-A MEDIAN IMPLANT PXL : 23X26 H 12MM 10 DEGREES
Common Device Name
ROI-A ALIF CAGE SYSTEM
Manufacturer (Section D)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
Manufacturer (Section G)
LDR MÉDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR   10300
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key14337382
Report Number3004788213-2022-00035
Device Sequence Number17708588
Product Code MAX
UDI-Device Identifier03662663011136
UDI-Public(01)03662663011136(17)240401(10)M55508
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K153495
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2024
Device Catalogue NumberIR2433P
Device Lot NumberM55508
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date05/09/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexFemale
-
-