Model Number 1804275-23
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat the left anterior descending (lad) coronary artery.The 2.75x23 mm xience skypoint stent was advanced to the target lesion; however the physician felt as though there was more balloon proximal to the marker on the stent and that the balloon went more proximal in the vessel than it should.The stent was implanted.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned device.The reported misassembled component could not be confirmed as the available evidence of the returned device demonstrated no manufacturing abnormalities.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.A cine was received and reviewed by an abbott vascular clinical specialist: the reviewer concluded that the images of the xience skypoint des 2.75 x 23 rx us provided confirmation that the device balloon, when inflated, did not extend to the end of the proximal stent edge.It appears the stent was fully expanded proximally despite the balloon appearing to be a couple of centimeters distal to the proximal stent edge.A device malfunction is visualized although a probable cause could not be determined.The investigation was unable to determine a conclusive cause for the reported misassembled component.In this case, due to analysis of the returned device and cine results, a conclusive cause for the reported misassembled component and noted failure to fold cannot be determined, as the reported device issue cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14337708
Report Number2024168-2022-05010
Device Sequence Number15033540
Product Code NIQ
UDI-Device Identifier08717648233173
UDI-Public08717648233173
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2023
Device Model Number1804275-23
Device Catalogue Number1804275-23
Device Lot Number1051741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/21/2022
Supplement Date Received by Manufacturer06/08/2022
Initial Report FDA Received Date05/09/2022
Supplement Report FDA Received Date06/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Patient Weight68 KG
-
-