It was reported that during a shoulder arthroscopy procedure, when passing the wire between the tendon, forceps of the 'first pass' broke, leaving the tip inside the joint.The doctor removed the tip from the joint using grasper forceps.The procedure was successfully completed with no surgical delay using a backup device.No further complications were reported.
|
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Based on the limited information provided the root cause of the reported device tip breakage could not be determined.Suture passer ifus outline precautionary statements and instructions regarding the use of the device to avoid damage or non-functionality, as misuse of this device may result in bent or broken distal tip or jaws.No patient injuries or adverse consequences were reported.No further medical assessment can be rendered at this time.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|