Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history: manufacturing location: (b)(4), manufacturing date: 13-aug-2021, expiration date: 31-jul-2030, part number: 04.001.224, 7mm ti cannulated humeral nail-ex/220mm-sterile, lot number: 323p992 (sterile), lot quantity: (b)(4).Production order traveler met all inspection acceptance criteria.Inspection sheet, in-process / inspect dimensional / final met all inspection acceptance criteria.Packaging label log was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn (b)(4) supplied by ethicon (b)(4) was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component(s) reviewed: component parts were not reviewed as the reported complaint condition of ¿removal due to various collapse of the bone and discomfort¿ does not indicate breakage of the nail or any of its components.Therefore, review of the raw materials would not be pertinent to the reported complaint condition.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|