| Model Number |
381244 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
04/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the needle was loose in the bd insyte¿ iv catheter during use.The following information was provided by the initial reporter, translated from (b)(6): "catheter without the part that fixes the bevel.When catheterizing the peripheral access in the patient, the beveled needle rotates.".
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: no photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.Based on the quality team's investigation, the root cause of this incident cannot be determined.Complaints received for this product and condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business regularly reviews the collected data for identification of emerging trends.
|
| |
|