Product complaint # (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: please provide procedure name.Cesarean section.Please clarify if there were any adverse patient consequences due to this event? no.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name - irgacare®, active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral, strength ¿ = 275 g/m.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Date sent to the fda: 8/9/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h6 h6 component code: g07002 no device problem found h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that a vcp357h sample was returned with the packet.Packet a contains a detached needle, winding former with a suture.Visual inspection was conducted on the returned device.After investigation of the needle and the suture of the sample a; revealed that clip off defect was observed in the sample.The extreme suture was noted to be severely cut.Some remnant suture could be observed through the barrel hole.As per returned conditions of the sample no functional test was performed.This defect of sample a is caused by excessive pressure on the suture during swaging of the needle and consequently breaking.Based on the information currently available, the clip-off was identified during the investigation of the sample received.This product issue will be addressed through ethicon inc¿s quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that two vcp357h samples were returned with two packets.The packet b & c contains a detached needle and a winding former.Regarding sample b & c, no conclusion could be reached on what may have caused the breakage suture since the suture was not received.In addition, the needles were examined, the swage and attachment area were noted to be as expected and no remnant suture was noted on the barre hole of the needles.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 g/m.
|