MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK SOLUTION ADMINISTRATION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8671
Medical Device Problem Code Free or Unrestricted Flow (2945)
Health Effect - Clinical Code High Blood Pressure/ Hypertension (1908)
Date of Event 04/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that during patient infusion of levophed, phenylephrine and amiodarone with 3 units of clearlink system extension sets, an inadvertent bolus from the extension filter was observed.This resulted in the patient experiencing hypertension reported as ¿caused the blood pressure to skyrocket¿.According to the reporter, the filter emptied out on its own.No connection issues were noted and the lines were setup properly.After the "filters were removed, the patient's blood pressure "evened out".No additional information is available.
 
Additional Manufacturer Narrative
The actual device was not available; however, a companion sample was received for evaluation.Visual inspection with the naked eye of the companion sample did not identify any abnormalities that could have contributed to the reported condition.Functional testing including clear passage and under water pressure testing were performed.The companion samples were determined to meet functional performance specification requirements.The reported condition was not verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK SOLUTION ADMINISTRATION SETS
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14339574
Report Number1416980-2022-02326
Device Sequence Number5337397
Product Code FPA
UDI-Device Identifier00085412046334
UDI-Public(01)00085412046334
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153158
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/06/2025
Device Catalogue Number2C8671
Device Lot NumberR20H06113
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/09/2022
Supplement Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMIODARONE; LEVOPHED; PHENYLEPHRINE
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-