ERIKA DE REYNOSA, S.A. DE C.V. SAFE LOCK APD LUER LOCK CONNECTOR; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Model Number SAFE LOCK APD LUER LOCK CONNECTOR
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the green clamp line during their pd treatment.It was reported that the connection may have been slightly loose.The patient reported receiving a supply bag lines are blocked message during dwell 5 of 5 of treatment.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).Upon follow up, the patient stated that they received an air detected in cassette message and supply bag lines are blocked message during dwell 5 of 5 and upon looking around to see what was causing the alarm the patient noticed that the baxter bag was disconnected and leaking.The patient stated that they were not sure why the baxter bag came disconnected.The patient stated that no fluid got in or near the cycler.The patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed, and confirmed that they were able to complete peritoneal dialysis treatment using continuous ambulatory peritoneal dialysis (capd) in the absence of the cycler.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cassette used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the green clamp line during their pd treatment.It was reported that the connection may have been slightly loose.The patient reported receiving a supply bag lines are blocked message during dwell 5 of 5 of treatment.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to discontinue the use of their cycler and follow up with their peritoneal dialysis nurse (pdrn).Upon follow up, the patient stated that they received an air detected in cassette message and supply bag lines are blocked message during dwell 5 of 5 and upon looking around to see what was causing the alarm the patient noticed that the baxter bag was disconnected and leaking.The patient stated that they were not sure why the baxter bag came disconnected.The patient stated that no fluid got in or near the cycler.The patient is trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed, and confirmed that they were able to complete peritoneal dialysis treatment using continuous ambulatory peritoneal dialysis (capd) in the absence of the cycler.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The cassette used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Additional Manufacturer Narrative
Corrections: d10 additional information: g1 plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE LOCK APD LUER LOCK CONNECTOR
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key14340564
Report Number0008030665-2022-00528
Device Sequence Number14752210
Product Code KDJ
UDI-Device Identifier00840861100811
UDI-Public00840861100811
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K124024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSAFE LOCK APD LUER LOCK CONNECTOR
Device Catalogue Number050-95018
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date05/09/2022
Supplement Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; LIBERTY CYCLER SET ; LIBERTY SELECT CYCLER; LIBERTY SELECT CYCLER
Patient SexUnknown
-
-