The lens was returned for evaluation.Solution is dried on the lens.The lens was cut in half, typical of insertion and removal from the eye.Scratches are observed on the optic.Cracks and marks are observed on the optic.Product history records were reviewed and the documentation indicated the product met release criteria.Associated products were not provided.It is unknown if qualified products were used.There have been no other complaints reported in the lot number the reported lens damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
|