HOLOGIC, INC. NOVASURE; DEVICE, THERMAL ABLATION, ENDOMETRIAL Back to Search Results
Model Number 2000
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.Device was returned: a visual inspection was performed and the device appeared to retain an overall acceptable integrity (including the sheath).No missing parts or missing components were detected.A functional test was performed, including cia, epa test, and ablation, and the device performed as expected.Hence the complaint cannot be confirmed.
 
Event or Problem Description
It was reported that a patient received a novasure procedure (b)(6) and that the procedure went well but after the procedure the md observed a piece of the array inside the uterus and removed it.The device was examined and it appears the metal came from it.No patient injury was reported.The device was returned and analized.Visual inspection was performed and the device appeared to reatin an overall acceptable integrity incluiding the sheath.No missing parts or missing componentes were detected.A functional test was performed including cia, epa test and ablation and the device performed as expected.Hence the complaint cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVASURE
Common Device Name
DEVICE, THERMAL ABLATION, ENDOMETRIAL
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough, MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jose, alajuela 20102- CRI
CS   20102 CRI
MDR Report Key14341586
Report Number1222780-2022-00125
Device Sequence Number5237229
Product Code MNB
UDI-Device Identifier15420045501140
UDI-Public(01)15420045501140(10)21G28RF(17)230218
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P010013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2023
Device Model Number2000
Device Catalogue NumberNS2000US
Device Lot Number21G28RF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Initial Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-