| Model Number |
MX-410410 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture and exchange from textured to smooth breast implants due to the patient¿s concern with the product.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side rupture and exchange from textured to smooth breast implants due to the patient¿s concern with the product.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: "analysis of the returned device identified: underweight and opening on posterior.A visual and microscopic analysis was performed which identified: sharp opening on posterior, wear abrasion, brown particles in the shell and crease fold.A dimension measurement in the shell was performed which identified the thickness within specification.Based on the device analysis the final assessment is: sharp opening on posterior assessed as an unidentified (tear) opening.".
|
| |
|