Model Number 050-50758
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
A peritoneal dialysis registered nurse (pdrn) reported when they performed a flush on the patient and was removing the disk from the organizer, they noted a fluid leak at the site where the drain line connects to the disk.The patient was disconnected but it was unknown if the leak was there from the start of the exchange or happened at the time the pdrn was removing the disk.The pdrn was advised to notify the physician of the event.The sample was discarded and was no longer available to be returned for evaluation.Additional information was requested but was not received to date.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A peritoneal dialysis registered nurse (pdrn) reported when they performed a flush on the patient and was removing the disk from the organizer, they noted a fluid leak at the site where the drain line connects to the disk.The patient was disconnected but it was unknown if the leak was there from the start of the exchange or happened at the time the pdrn was removing the disk.The pdrn was advised to notify the physician of the event.The sample was discarded and was no longer available to be returned for evaluation.Additional information was requested but was not received to date.
 
Additional Manufacturer Narrative
Upon review of the available information, it was determined that the event reported in mfr (b)(4) was related to a fluid leak from a drain line after treatment, and is not a reportable event.There was no serious injury, adverse event, or reportable malfunction.There will be no further updates on (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STAY SAFE MTS SET
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key14342618
Report Number0008030665-2022-00533
Device Sequence Number4961288
Product Code KDI
UDI-Device Identifier00840861100699
UDI-Public00840861100699
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K904806
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number050-50758
Device Catalogue Number050-50758
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/04/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date05/09/2022
Supplement Report FDA Received Date06/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-