MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO FLAT MARKERS; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO FLAT MARKERS; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSDV0016
Medical Device Problem Codes Material Twisted/Bent (2981); Intermittent Communication Failure (4038)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during a cori assisted tka surgery, the real intelligence femur tracker was blinking in and out of camera view despite the camera was looking directly at it.Also, an excessive amount of writing was noticed on the reflective surface of the flat marker; therefore, they swapped the noted flat marker out, but it did not help.The procedure was resumed, after a non-significant delay, by changing to manual procedure.No patient complications were reported.Moreover, the real intelligence femur tracker appears to be bent.
 
Additional Manufacturer Narrative
H3, h6: the real intelligence femur tracker, part number rob10018, lot number 20bct0031, used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The real intelligence cori for knee arthroplasty user manual (500230 rev d) provides guidelines for recovering to a fully manual procedure in the "recovery procedure guidelines, appendix d".The failure mode and associated risk have been anticipated within the risk file.The risk level is still adequate.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with a loose flat marker connection to the femur tracker.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO FLAT MARKERS
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14342729
Report Number3010266064-2022-00351
Device Sequence Number17463978
Product Code OLO
UDI-Device Identifier00885556628904
UDI-Public00885556628904
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K201022
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberPFSDV0016
Device Catalogue NumberPFSDV0016
Device Lot Number22BK00194
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer08/04/2022
Initial Report FDA Received Date05/09/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-