| Model Number |
60PFS50 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that whilst in use in a young person, the eyelet was damaged, close to fully splitting whilst on the tube's 4th use.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during this device history record review.A product sample was received for evaluation.Visual and functional testing were performed.Sample was visual inspection under normal conditions of illumination to detect any defect.A broken eyelet was found.The root cause of the reported issue was found to be the eyelet was damaged.Actions were taken to mitigate the reported issue: notification was made to production personnel to explain the importance to adherence or following in the manufacturing procedure.
|
| |
|