It was reported by the olympus representative on behalf of the customer, fluid invaded the uretero-reno fiberscope.The issue was found at reprocessing after a therapeutic procedure.The procedure was not affected.No patient harm reported.During the evaluation of the device, it was noted the bending section had a protruding broken skeleton.This report is to capture the reportable malfunction of the bending section had a protruding broken skeleton noted at estimation.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record (dhr) was unable to be reviewed for this device however, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the following likely caused the event.The user applied excessive stress to bending section, causing breakage of the bending tube.The root cause of why the device was not reprocessed according to the device instructions for use (ifu) could not be determined.The following is included in the ifu and may have helped prevent the event: "perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk." "inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities." "do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist." "do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged." olympus will continue to monitor field performance for this device.
|