(b)(4).Date sent: 08/29/2022.D4: batch # t5d14h.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the ecs21a device arrived with no apparent damage.The breakaway washer was present, cut and there was no staples present, indicating that the device achieved a full firing stroke.In addition, the washer was noted to have nicks; this damaged is consistent when the device is fired over an already existing staple line, hard object or thicker tissue than indicated.The device was reloaded with staples, a new washer was placed on the device and it was tested for functionality.It fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: the firing stroke must be completed.Do not partially fire the instrument.Incomplete firing can result in malformed staples, incomplete cut line, bleeding, and leakage from the staple line and/or difficulty removing the device.Ensure that the firing trigger is fully squeezed to ensure proper staple formation and cutting of tissue.The event described could not be confirmed as the device cut and performed without any difficulties.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|