| Model Number |
CSE-P-125 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hypovolemia (2243)
|
| Date of Event |
02/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A haemonetics field service engineer evaluated the data log files from the cell saver elite system used in the procedure.Data showed that the device functioned correctly, tripping to wash @ the correct point.There is no evidence to suggest that the event was related to the device used during the procedure.There was no report of patient adverse reaction in this event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2022, haemonetics was notified of an early wash trigger with an estimated blood loss approximately 1500 ml (pre-op hb/ 97g/dl); cell salvage blood re-infused approximately 114 ml during a procedure using a cell saver® elite® plus autotransfusion system and cell saver® elite set-125m in the (b)(6).There was no report of patient injury.
|
| |
|