Model Number A34-34/C100V
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/02/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device remains implanted in the patient; therefore, it will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially implanted with a afx bifurcated stent graft and an afx vela infrarenal stent graft extension to treat an abdominal aortic aneurysm (aaa).Approximately, 1 month post initial procedure a computed tomography (ct) identified a 1a endoleak.Physician elected to treat the patient with a medtronic (non-endologix) aortic cuff.Note: this event is for the type 1a endoleak re-intervention which was discovered during a recent reported event for a type 3a endoleak with separation of the vela infrarenal and the medtronic (non-endologix) aortic cuff.The event for the 3a endoleak will be reported under a separate report.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type 1a endoleak and additional endovascular procedure are confirmed.This is consistent with the reported adverse.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was not reported.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially implanted with a afx bifurcated stent graft and an afx vela infrarenal stent graft extension to treat an abdominal aortic aneurysm (aaa).Approximately, 1 month post initial procedure a computed tomography (ct) identified a 1a endoleak.Physician elected to treat the patient with a medtronic (non-endologix) aortic cuff.Note: this event is for the type 1a endoleak re-intervention which was discovered during a recent reported event for a type 3a endoleak with separation of the vela infrarenal and the medtronic (non-endologix) aortic cuff.The event for the 3a endoleak will be reported under a separate report (mfr.Report # 2031527-2022-00098).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
VELA INFRARENAL
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key14347216
Report Number2031527-2022-00097
Device Sequence Number20151205
Product Code MIH
UDI-Device Identifier00818009013552
UDI-Public(01)00818009013552(17)190906
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/06/2019
Device Model NumberA34-34/C100V
Device Lot Number1587102008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/28/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date05/10/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/21/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX, BIFURCATED STENT GRAFT, LOT#1283281001
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
-
-