• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551NAS
Device Problems Break (1069); Device Difficult to Program or Calibrate (1496); Priming Problem (4040)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/04/2022
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the insulin pump had keypad anomaly.Customer stated the insulin pump would get stuck sometimes and arrow up and down buttons sometimes stick.Customer stated that there was cracked around the reservoir area.No harm requiring medical intervention was reported.The device will not be retuned for analysis.
 
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Additional information has been added which was not included with the initial report.The information has been provided in section b5 with this report.On the event date (b)(6) 2022.Customer returned pump for an alleged prime/fill anomaly and intermittent button response.Pump passed the displacement test, selftest, seat, rewind, basic occlusion test, occlusion test, prime/a33 test and excessive no delivery test.All button functioning properly noted.Pump history download using thds was successful.However, a47 alarms found on the history file which caused corrupted data.Unable to confirm the reported event date due to the corrupted data.A47 alarms are due to the pump not having power for a long period of time.Unit was cut/open and inspected no moisture damage or component damage found inside the pump.Test reservoir lock properly inside the reservoir tube.The following were noted during visual inspection: broken battery tube threads and cracked reservoir tube lip.Unit passed required testing.Unit functioning properly during testing.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Occupation has been updated and provided with this report in section e3.
 
Event Description
The device was returned for the analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key14347413
MDR Text Key299878157
Report Number2032227-2022-188802
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000643169404298
UDI-Public(01)000000643169404298
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-551NAS
Device Catalogue NumberMMT-551NAS
Device Lot NumberA2551NASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 05/05/2022
Initial Date FDA Received05/10/2022
Supplement Dates Manufacturer Received01/06/2023
Supplement Dates FDA Received01/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age40 YR
Patient SexMale
Patient Weight84 KG
-
-