• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS VOLCANO EAGLE EYE PLATINUM ST; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS VOLCANO EAGLE EYE PLATINUM ST; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number 85900PST
Medical Device Problem Codes Entrapment of Device (1212); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This case was reviewed and investigated according to the manufacturer¿s policy.Weight: no information available.Relevant tests/history: no information available.Implant/explant date: not applicable for this device.The eagle eye platinum st catheter was not returned for evaluation, thus no returned product investigation was performed.Recall: do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
It was reported that during a diagnostic peripheral procedure, resistance was encountered during insertion of the eagle eye platinum st catheter within the upper right arm extremity.When the physician attempted to pullback the catheter, more resistance was encountered.Ultimately, the catheter sheered and the distal tip separated from the catheter.The proximal portion of the catheter was removed and the 0.014¿ non-manufacturer guide wire was bent.The separated tip was removed with a snare and a full body fluoroscopy confirmed no piece was retained inside the patient.The patient was discharged as expected in good condition.This adverse event and product problem is being submitted due to the tip separation requiring intervention (snare).
 
Additional Manufacturer Narrative
Block h3: the eagle eye platinum st catheter was returned in two pieces.The catheter separated at the distal shaft approximately 185 mm from the tip (includes distal tip, distal and proximal fillets, scanner body, and a portion of the distal shaft).Under visual and microscopic inspection, the microcables and inner member were exposed with rough edges of malleable material.Additionally, the core wire was exposed with a sharp edge.There was no missing material, all portions of the catheter was accounted for.Block h6 correction: component code updated from 3123 (tip) to 4721 (rod/shaft).Based on the device evaluation, the distal shaft separated and not the distal tip.Block h6: the probable cause of the shaft separation is damage during use and handling.Strain, impact, and forces associated with use can affect the integrity of the device.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EAGLE EYE PLATINUM ST
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
PHILIPS VOLCANO
3721 valley centre drive #500
san diego CA 92130
Manufacturer (Section G)
VOLCARICA S.R.L.
coyol free zone &business park
b37
alajuela
CS  
Manufacturer Contact
ayse yarimoglu
3721 valley centre drive #500
san diego, CA 92130
MDR Report Key14347478
Report Number3008363989-2022-00027
Device Sequence Number12014352
Product Code OBJ
UDI-Device Identifier00845225002312
UDI-Public(01)00845225002312(11)211213(17)231213(10)0302477134
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K120697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/13/2023
Device Model Number85900PST
Device Catalogue Number400-0200.141
Device Lot Number0302477134
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date05/10/2022
Supplement Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASAHI: 0.014 GLADIUS GUIDEWIRE; TERUMO: 6FR DESTINATION INTRODUCER SHEATH
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexMale
Patient EthnicityHispanic
Patient RaceWhite
-
-