• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE BURETTE INFUSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS PUMP MODULE SMARTSITE BURETTE INFUSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 2447-0007
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd alaris pump module smartsite burette infusion set was clogged.The following information was provided by the initial reporter: trying to prime ivig (intravenous immune globulin) and noticed that the medication wasn't going into the ball chamber in order to prime.Squeezed it multiple times, flipped it around, opened all the clamps, but it wasn't going into the ball chamber.Had to remove the ivig in the burette since it wouldn't prime.No harm came to the patient, but luckily, we were able to pull the medication out of the chamber with the port attached.
 
Additional Manufacturer Narrative
Date of birth: only the patient's age was provided, therefore a default date of birth has been listed.Fda notified: the initial reporter notified the fda via uf/importer report#: (b)(4).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H6: investigation summary no product or photo was returned by the customer.The customer complaint that the product won't prime could not be verified due to the product not being returned for failure investigation.A device history record review for model 2447-0007 lot number 21095293 was performed.The search showed that a total of (b)(6) in 1 lot number was built on 09sep2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed and a root cause could not be determined.
 
Event or Problem Description
It was reported that the bd alaris pump module smartsite burette infusion set was clogged.The following information was provided by the initial reporter: trying to prime ivig (intravenous immune globulin) and noticed that the medication wasn't going into the ball chamber in order to prime.Squeezed it multiple times, flipped it around, opened all the clamps, but it wasn't going into the ball chamber.Had to remove the ivig in the burette since it wouldn't prime.No harm came to the patient, but luckily, we were able to pull the medication out of the chamber with the port attached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS PUMP MODULE SMARTSITE BURETTE INFUSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14347485
Report Number9616066-2022-00568
Device Sequence Number5009995
Product Code FPA
UDI-Device Identifier50885403235076
UDI-Public50885403235076
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K944320
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/09/2024
Device Model Number2447-0007
Device Catalogue Number2447-0007
Device Lot Number21095293
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Supplement Date Received by Manufacturer05/10/2022
Initial Report FDA Received Date05/10/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age16 YR
Patient SexMale
-
-