| Model Number |
HST III SEAL (4.5MM), 5-PACK |
| Medical Device Problem Code |
Fitting Problem (2183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The hospital reported that at the time of opening hst iii seal (4.5mm), 5-pack, the seal was not in the normal position and it was not ready for loading, so they did not use it for the patient and finished the surgery with a new product.The seal did not fit properly after opening.There was no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise # (b)(4).Analysis of production: (3331/213/67) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis for similar complaints: (4109/213/67) there were one additional similar complaint(s) reported for the reported failure mode and the reported lot number.A review of the product complaint trend data was performed for the months off ¿dec 2021¿ through ¿mar 2022¿.The review does not identify an increasing trend or an adverse trend for three consecutive months for the product family and the failure mode.Trend analysis: (4110/213/67) the overall 24 month product complaint trend data for the period apr-2020 through mar-2022 was reviewed.There were no triggers identified for the review period.Communication/interviews: (4111/213/67) communication/interviews were performed to obtain all possible information.
|
| |
|
Search Alerts/Recalls
|